- Products
Helium Liquefier & Recovery System
Helium liquefy
KDHR15 Helium Reliquefier
Liquefaction rate | 15L/d@1 PSI, 20L/d@5PSI | ||
Reliquefaction rate | ≥20L/d@1 PSI | ||
Vibration index | ±2 μm | ||
Compressor power supply | 3 phase, 380V (50Hz)/480V(60Hz) | ||
Control unit power supply | Single Phase AC110V-240V (50-60Hz) | ||
Cooling water | Inlet temperature: 5~25℃; Flow rate 7~10L/min; Pressure<8bar | ||
Helium Purity requirement | >99.999% | ||
Liquid helium infusion line | Material:sus304, O.D: 9.5mm, Lenth:550 mm(can be customized) | ||
I-Liquefier20
Liquefaction rate | 20L/d | ||
Dimension | ≥1600x700x1500mm (compressor included) | ||
VWeight | 330kg | ||
Cold head | KDE415SA | ||
Compressor | KDC6000V | ||
Quantity of cryocooler | 1 set | ||
Dewar | 150L | ||
Cooling time | <4h | ||
Power supply of control unit | Single PH AC110V—240V 50~60Hz (Power consumption<1KW) | ||
Cooling water | Inlet temperature 5~25℃, flow rate7-9L/min, pressure<8bar | ||
I-Liquefier 40
Liquefaction rate | 40L/d | ||
Dimension | 1650x760x1735mm (compressor included) | ||
VWeight | 600kg | ||
Cold head | KDE415SA | ||
Compressor | KDC6000V | ||
Quantity of cryocooler | 2 set | ||
Dewar | 200L | ||
Cooling time | <4h | ||
Power supply of control unit | Single PH AC110V—240V 50~60Hz (Power consumption<1KW) | ||
Cooling water | Inlet temperature 5~25℃, flow rate7-9L/min, pressure<8bar | ||
KDHRR80 Helium Liquefier
Liquefaction rate | 72L/day(5PSI) 80L/day(10PSI) | ||
Dimension | unsure | ||
VWeight | unsure | ||
Cold head | KDE415SA | ||
Compressor | KDC6000V | ||
Quantity of cryocooler | 4 set | ||
Dewar | 500L(can be selected according to customer requirements) | ||
Cooling time | <4h | ||
Power supply of control unit | 3PH AC380V 50Hz (Power consumption: 28~32KW)/ 3PH AC380V 60Hz (Power consumption: 28~32KW) | ||
Cooling water | Inlet temperature 5~25℃, flow rate28~36 L/min, pressure<8bar; Pure water | ||
HDHRR100 Helium Liquefier
Liquefaction rate | 90L/day(5PSI) 100L/day(10PSI) | ||
Dimension | unsure | ||
VWeight | unsure | ||
Cold head | KDE415SA | ||
Compressor | KDC6000V | ||
Quantity of cryocooler | 5 set | ||
Dewar | 1000L(can be selected according to customer requirements) | ||
Cooling time | <4h | ||
Power supply of control unit | 3PH AC380V 50Hz (Power consumption: 35~40KW)/ 3PH AC380V 60Hz (Power consumption: 35~40KW) | ||
Cooling water | Inlet temperature 5~25℃, flow rate28~36 L/min, pressure<8bar; Pure water | ||
KDE210SA
Liquefaction rate | 20L/d | ||
Dimension | ≥1600x700x1500mm (compressor included) | ||
VWeight | 330kg | ||
Cold head | KDE415SA | ||
Compressor | KDC6000V | ||
Quantity of cryocooler | 1 set | ||
Dewar | 150L | ||
Cooling time | <4h | ||
Power supply of control unit | Single PH AC110V—240V 50~60Hz (Power consumption<1KW) | ||
Cooling water | Inlet temperature 5~25℃, flow rate7-9L/min, pressure<8bar | ||
KDE400SX
Coldhead Type | KDE400SX | ||
Lowest Temperature: | < 3.5K | ||
Cooling Capacity (50Hz): | First Stage | Second Stage | |
35W @ 50K | 1.7W @ 4.2K | ||
Cooldown Time (2nd stage): | < 60min.(4.2K) | ||
Weight: | Coolhead | Compressor | |
19 kg | 118 kg | ||
Compressor Type: | KDC6000V | ||
Power Consumption(50Hz): | Steady State | Cooldown | |
6.5kW | 7.2kW | ||
Compressor Cooling Requirement: | >7 L/Min Water | ||
Mainteance: | >10.000 Hours |